VALIDITY OF WEARABLE TECHNOLOGY TO MEASURE PEAK IMPACT DURING HIGH-INTENSITY TREADMILL RUNNING

##article.authors##

  • Daniel Wundersitz
  • Paul Gastin
  • Chris Richter
  • Kevin Netto

##semicolon##

Motion analysis##common.commaListSeparator## accelerometer##common.commaListSeparator## acceleration##common.commaListSeparator## collision

##article.abstract##

The purpose of this study was to identify the validity of an upper-body mounted accelerometer to measure peak acceleration during high-intensity treadmill running. A twelve camera motion analysis (MA) system was used as the criterion measure with markers placed on and close to the accelerometer. Ten peak impacts per participant were compared (n = 390). All accelerometer values were significantly different between the MA unit and T6 reflective marker’s acceleration data. Smoothing accelerometer data at 8 and 6 Hz provides an acceptable indirect measure of peak impact acceleration performed during high-intensity running. Consequently, smoothing algorithms should be incorporated into the commercially available software that the devices are supplied with.

##submission.downloads##

##submissions.published##

2013-09-01

##issue.issue##

##section.section##

Equipment / Instrumentation